A four-electrode microconstant direct current resistance detector for ion chromatography applying ion-exchange membrane and porous electrode

Release time:2020-01-01    Hits:

Affiliation of Author(s):Xiamen University

Journal:Talanta

Key Words:Resistance detector; Ion chromatography; Four-electrode; Porous electrode; Constant current

Abstract:A four-electrode microconstant direct current resistance detector for ion chromatography not sensitive to the effects of electrode polarization, capacitance, and electrolysis by-products is proposed. A constant current of microampere magnitude is applied across the current electrodes of the four-electrode device, and the voltage responses between the detection probes are directly picked up by a high input impedance instrumentation amplifier. The ion-exchange membranes, which separate the detection chamber from the electrolysis chambers, enable the measurement of solution resistance free of the

Co-author:Weixiong Huang, Huabin Chen, Yuhua Su, Rongzong Hu

First Author:Weixiong Huang

Indexed by:Journal paper

Correspondence Author:Rongzong Hu

Document Code:10.1016/j.talanta.2010.07.003

Discipline:Science

First-Level Discipline:Chemistry

Volume:82

Page Number:1364-1370

Translation or Not:no

Date of Publication:2010-08-04

Included Journals:SCI

Links to published journals:https://www.sciencedirect.com/science/article/pii/S0039914010005175