Doctoral Supervisor
The Last Update Time: ..
Release time:2020-01-01 Hits:
Affiliation of Author(s):Xiamen University
Journal:Talanta
Key Words:Resistance detector; Ion chromatography; Four-electrode; Porous electrode; Constant current
Abstract:A four-electrode microconstant direct current resistance detector for ion chromatography not sensitive to the effects of electrode polarization, capacitance, and electrolysis by-products is proposed. A constant current of microampere magnitude is applied across the current electrodes of the four-electrode device, and the voltage responses between the detection probes are directly picked up by a high input impedance instrumentation amplifier. The ion-exchange membranes, which separate the detection chamber from the electrolysis chambers, enable the measurement of solution resistance free of the
Co-author:Weixiong Huang, Huabin Chen, Yuhua Su, Rongzong Hu
First Author:Weixiong Huang
Indexed by:Journal paper
Correspondence Author:Rongzong Hu
Document Code:10.1016/j.talanta.2010.07.003
Discipline:Science
First-Level Discipline:Chemistry
Volume:82
Page Number:1364-1370
Translation or Not:no
Date of Publication:2010-08-04
Included Journals:SCI
Links to published journals:https://www.sciencedirect.com/science/article/pii/S0039914010005175
AttachmentsA four-electrode microconstant direct current resistance detector for ion chromatography applying ion-exchange membrane and porous electrode.pdf Download: [] times