A novel anti-interference and pH-modulation device: application in enzyme-free glucose detection

点击次数:

发布时间:2020-01-01

所属单位:Xiamen University

发表刊物:Microchim Acta

关键字:Anti-interference; pH-modulation; Electrodialysis; Glucose; FIA-amperometric detection

摘要:We report on a novel anti-interference and pHmodulation device (herein after referred to as "device"). It is based on electrodialysis and can continuously increase the pH value of the carrier solution and - at the same time - remove interfering analytical signals obtained for ascorbic acid (AA) and uric acid (UA). The "device" was coupled to the FIA amperometric detection of glucose. The linear range is from 1 μmol L−1 to 0.4 mmol L−1, with a sensitivity of 213 μA cm−2 mM−1 and a detection limit of 1 μmol L−1 at a signal-to-noise ratio of 3. The method was used to sucessfully determine glucose

合写作者:Yuhua Su, Rongzong Hu, Weixiong Huang, Kangkang Hu