Characterization of ion exchange functionalized cyclic olefin polymer open tubular columns
点击次数:
发布时间:2020-01-01
所属单位:The University of Texas at Arlington
发表刊物:Analytica Chimica Acta
关键字:Column capacity; Column uniformity; Cycloolefin polymer; On-column titration; Admittance detection
摘要:We describe both the cation-exchange (CEX) and anion-exchange (AEX) capacity measurements of functionalized cyclic olefin polymer (COP) open tubular (OT) columns. COP capillaries were sulfonated to provide CEX functionality and then coated with AEX latex to provide AEX functionality. We measure functionalization uniformity along the column by (a) separately measuring capacities of two halves of a column, (b) measuring retention factors for several ions in successive 10 cm segments of a column in multiple injections and (c) measuring retention variance of a single ion in different segments of a
合写作者:Weixiong Huang, Sasikarn Seetasang, Purnendu K. Dasgupta
第一作者:Weixiong Huang
论文类型:期刊论文
通讯作者:Purnendu K. Dasgupta
论文编号:10.1016/j.aca.2018.06.063
学科门类:理学
一级学科:化学
卷号:1036
页面范围:187-194
是否译文:否
发表时间:2018-06-26
收录刊物:SCI
发布期刊链接:https://www.sciencedirect.com/science/article/pii/S0003267018308286?via%3Dihub
附件
Characterization of ion exchange functionalized cyclic olefin polymer open tubular columns.pdf